FUJIFILM Sonosite, Inc. ("Sonosite") is aware of the recent BlueBorne vulnerability and has diligently evaluated the potential impact on its ultrasound systems and operating software. After a thorough investigation, Sonosite has determined that the systems listed below are not affected by BlueBorne:

  • Sonosite Edge II ultrasound system
  • Sonosite SII ultrasound system
  • Sonosite X-Porte ultrasound system
  • Sonosite Edge ultrasound system
  • Sonosite NanoMaxx ultrasound system
  • Sonosite M-Turbo ultrasound system
  • Sonosite S Series ultrasound systems
  • Sonosite MicroMaxx ultrasound system
  • Sonosite Titan ultrasound system
  • Sonosite iLook ultrasound system
  • Sonosite 180 ultrasound system
  • Sonosite SonoHeart ultrasound system
  • FUJIFILM FC1 ultrasound system

The following system is shipped in a state that mitigates the BlueBorne vulnerability:

  • Sonosite iViz ultrasound system

The Sonosite iViz product has Bluetooth support present on the system but is shipped with Bluetooth disabled (default). Customers should be advised to leave the Bluetooth configuration of the product in its as-shipped state, which is disabled. Disabling Bluetooth mitigates the vulnerability.

Sonosite remains committed to the safety and security of its customers, any further questions can be directed to Sonosite technical support .